Support Hub

Welcome to the Bio Molecular Systems Support Hub for Myra and Mic products.

Frequently Asked Questions

See if we can answer your question here. If not please proceed below to create your support enquiry.

Why do some of my Mic runs have very low fluorescence levels?

A common cause of unexpectedly low fluorescence is leaving old reaction tubes in the rotor from a previous run. The Mic automatically sets its signal gain based on the tube with the highest fluorescence. If leftover tubes contain a stronger residual signal than your new samples, the instrument will scale the gain too low, making your current reactions appear weak.

This is not an assay or reagent issue but a setup artifact.

To avoid this, always remove tubes after each run. Alternatively, use consistent “water load” tubes or manually select a tube for gain adjustment to ensure accurate signal scaling.

[Download Resource]

Do I need to balance the Mic rotor?

Yes. While the Mic rotor does not require mechanical balancing like a centrifuge, maintaining a consistent thermal load across all positions is essential for accurate results. Empty tubes, missing tubes, or differing reaction volumes can create thermal gradients in the rotor, leading to increased variability between samples.

To ensure optimal performance, always fill every rotor position, use water-filled tubes for unused positions at the same volume as your reactions, and verify that all tubes are properly seated and secured.

Can I run the Mic on alternative power sources (e.g., in the field)?

Yes. The Mic is designed to operate reliably in field conditions using alternative power sources such as batteries, solar setups, or generators. When running from a battery or solar system, you will need a pure sine wave inverter with at least 360 W continuous output. Square wave inverters are incompatible.

For unstable power sources (e.g., generators), using a pure sine wave online UPS is recommended to provide clean, consistent power and prevent interruptions during a run. With the correct setup, the Mic can perform reliably outside of standard laboratory environments.

[Download Resource]

Why does the Mic never need calibration?

The Mic is designed so that calibration is not required. Its induction-based heating system directly heats the aluminum rotor, eliminating the temperature drift commonly seen in block-based cyclers. As a result, temperature accuracy remains stable over time when the instrument is used as recommended.

In addition, the Mic has a fixed optical path with no moving parts, so there is no need for optical alignment, crosstalk compensation, or ongoing optical calibration.

Because of this stable design, the Mic does not require routine calibration and is supported by a 3-year factory warranty covering materials and workmanship.

[Download Resource]

How do I use the Mic with a LIMS?

The Mic integrates with LIMS via simple file-based import and export using standard formats such as .csv, .tsv, .txt, and .xlsx.

You can import sample and assay information from your LIMS into Workbench to set up a run, then automatically export run data and results back to your LIMS for storage and reporting. Workbench also supports custom metadata fields, allowing you to include LIMS-specific identifiers and tracking information with each run.

For streamlined workflows, results can be automatically saved or exported to a designated folder linked to your LIMS, with options to control export behavior (e.g., only export if no errors are detected).

Can I perform a protein thermal shift assay on the Mic?

Yes. The Mic is great for protein thermal shift assays because it can control temperature very accurately and has a high-resolution melting feature.

The instrument’s small, low-mass rotating aluminum rotor allows for very quick thermal equilibration and very little temperature difference between samples. This results in highly uniform and reproducible temperature ramping, which is critical for accurately determining protein melting temperatures (Tm). The Mic’s high data acquisition density and low signal-to-noise ratio during melt also make it possible to see unfolding transitions in great detail, which is especially useful for finding subtle stability shifts.

For DSF experiments, you can use dyes that increase fluorescence when binding to hydrophobic regions exposed during protein unfolding, as long as they are compatible with the Mic’s fixed optical channels. Examples include the GloMelt Thermal Shift Protein Stability Kit (green channel, validated by users on Mic). The assay’s suitability will depend on how well the dye’s excitation and emission profile matches the Mic channels that are available, as well as the protein’s expected melting range and buffer conditions.

Why is ROX used in qPCR, and why isn’t it needed on the Mic?

ROX is a passive reference dye used to normalize optical variability between wells in many block-based qPCR systems.

It primarily compensates for:

  • Differences in optical path length
  • Well-to-well variation in illumination and detection
  • Minor pipetting differences affecting fluorescence intensity

The Mic eliminates these sources of variability through its design:

  • All samples are measured in the same optical position
  • Rotor-based detection removes well-to-well optical differences
  • Highly consistent reaction geometry and volume

As a result, fluorescence signals are inherently uniform, and ROX normalization is not required.

What are the routine maintenance requirements for the Mic?

The Mic is designed to require no routine maintenance.

Unlike traditional qPCR systems, the Mic has no moving optical components, no need for calibration, and no consumable parts that require regular replacement. Its sealed optical path and solid-state temperature design minimize wear and eliminate the need for periodic servicing.

While no maintenance is required, basic care will help ensure optimal performance:

  • Keep the instrument clean and free of dust or spills
  • Ensure tubes and caps are properly seated before each run
  • Avoid introducing condensation or contaminants into the rotor

No scheduled servicing or calibration is required, just standard excellent lab practice.

Why do I see a “Driver_Supply_Voltage” error when starting my Mic?

This error indicates that the Mic is not receiving stable or sufficient power. It is typically related to the external power supply or mains input rather than the instrument itself.

Common causes include:

  • Unstable mains power (e.g., voltage dips or fluctuations)
  • Changes to power setup (different wall socket, circuit, or extension)
  • Incompatible or underperforming UPS/inverter systems
  • Faulty power adapter (power brick)

How to troubleshoot:

  • Try a different wall socket or circuit
  • Revert any recent changes to your power setup
  • If using a UPS or inverter, ensure it meets Mic requirements (e.g., sufficient output, stable supply)
  • If available, swap the power adapter with a known working unit to test

If no clear cause is identified, the power adapter may be faulty. Please contact your local distributor (if under warranty) or BMS support with details of your setup, including any UPS or inverter being used.

Why is my analysis showing a “Potential Optical Failure” warning?

This warning is triggered when little or no fluorescence signal is detected in a channel during a run.

In most cases, the warning indicates that no measurable signal was generated in the reaction, rather than an instrument fault.

Common causes include:

  • No fluorescent chemistry present (e.g., dry run or missing dye/probe)
  • Reaction setup issues (e.g., incorrect mix, missing components)
  • Reagent problems (e.g., degradation due to improper storage)
  • Incorrect channel selection for the assay

What to check:

  • Confirm the correct fluorescent dye/probe and channel are being used
  • Verify reaction setup and components
  • Check reagent integrity and storage conditions

This warning usually means fluorescence was not produced, indicating the optics are functioning.

What should I do if I see “Insufficient User Privileges” when opening Workbench?

This message usually appears when user permissions were enabled during installation, but the required Windows user groups have not been configured on the computer.

If you have administrator access:

  • Right-click the Workbench application and select “Run as administrator.”
  • Once open, go to the settings via the “?” menu icon and disable user permissions

If you do not have administrator access:

  • Uninstall Workbench
  • Reinstall the software and ensure the “Enable user permissions” option is not selected during installation

How can I transport the Mic?

For the safest transport, use the original packaging. Ensure the rotor is empty and all accessories are stored in their designated foam slots.

For fieldwork, a Pelican case with custom foam inserts designed to snugly fit the Mic, power adapter, cables, tubes, and accessories is recommended.

Why is my website registration not working?

Most registration issues involve the activation email. After submitting your form, you will receive an automated link to complete the process. If it does not appear in your inbox, please check your spam or junk folder. If you still cannot activate your account, contact support for assistance.

What can cause a drip to occur from my Myra?

The most common cause of a drip occurring during a Myra run is incorrect plate height calibration.

Incorrect plate height calibration can result in the following:

  • The tip enters the well too far, thereby collecting excess liquid on its outer surface.
  • The tip is touching the bottom of the well, so when the tip retracts, the resulting vacuum causes the liquid at the bottom to splash upward, resulting in excess liquid being retained on the tip.

Always use the correct consumable type indicated in the software and ensure calibration has been conducted correctly.

How do I change the bucket fill orientation from horizontal to vertical?

Use the space bar to toggle the fill order from horizontal to vertical. Ensure the bucket fill operation is active to enable the change.

How can I transport the Myra?

We recommend only transporting Myra in its original packaging.

To prepare the instrument, use the Lock Myra for Transport wizard in the Workbench software. You can find the lock by clicking the instrument icon, selecting Properties, and following the on-screen prompts to secure it safely.

What are the computer requirements for the software?

  • Windows 10, 64-bit (English version) operating system
  • Intel Core i5, Core 3, Core Ultra 3, or equivalent AMD processor (ARM CPUs like the
  • Snapdragon family are not supported)
  • 8 GB RAM (16 GB recommended)
  • 3 GB free hard drive capacity
  • 1080p screen resolution
  • One USB port
  • Adobe Reader must be installed to view reports in PDF format

The computer requirements are also listed at the start of the user manual.

How do I use multi-dispense?

From Workbench version 2.0.7 onwards, there is a checkbox available to turn multi-dispense on.

This can be found in the Information section under the Advanced Settings options.
For MyraScripts, a multi-dispense parameter will be coded in as a configurable option.

Which NGS Library Prep kits can I run on the Myra?

A comprehensive number of NGS library prep kits can be run on the Myra. Some of the more popular ones include:

  • InfiniSeq Myra Protocol
  • 10X Genomics Chromium Next GEM Single Cell Multiome ATAC plus Gene Expression, Step 3Post GEM Incubation Cleanup
  • seqWell plexWell LP384
  • NEBNext Ultra II FS DNA Library Prep Kit for Illumina
  • Zymo Quick-16S Plus NGS Library Prep Kit
  • Illumina DNA Prep
  • ONT Ligation Sequencing Kit V14
  • ONT Rapid Barcoding Kit 24/96 V14
  • ONT Native Barcoding Kit (24) V14
  • PacBio 16S Barcoding

You can find a list of NGS library preparation kits that have been validated for use on the Myra at the NGS library preparation applications page. More NGS library preparation kits are under development, so please contact us if you are interested in testing any library preparation kit that is not yet fully validated.

Where can I buy Myra tips?

You can buy Myra tips from your local BMS distributor or from our direct sales teams in the United States and the United Kingdom.

What are the routine maintenance requirements for Myra?

The Myra has minimal routine maintenance requirements, with only the HEPA filter replacement and calibration of the pipette heads as recommended tasks. The frequency of these maintenance tasks is determined by your laboratory and regulatory requirements.

Do I need to calibrate my plates and tube positions periodically?

Due to its use of optical encoders, it is not necessary to periodically calibrate the position of consumables once they have been initially calibrated. It remains possible to recalibrate items whenever required.

How do I clean or decontaminate my Myra?

Surfaces of the instrument, including the deck, can be decontaminated using a solution of sodium hypochlorite (NaOCl).

  • A solution containing 1 gL⁻¹ available chlorine is suitable for general laboratory sanitation
  • Stronger solutions of 5 gL⁻¹ are recommended for high-risk situations (WHO, 2020)

Alternatively, reagents such as DNA Away or milder decontamination solutions can also be used.

Avoid applying these reagents to exposed electrical components such as the back of the pipette head or axis motors.

What happens if MyraVision can’t recognize my consumable?

If MyraVision cannot recognize a consumable that it indicates it should be able to recognize, confirm that the consumable is correct and positioned correctly, then ignore the error and start the run. Following the run, create a support package and send it to the support team.

What can cause an issue with pipetting during a Myra run?

Common causes of pipetting issues during a Myra run include the following:

An incorrect consumable was used instead of the one selected in the software.

Use consumables in the software that exactly match what is placed on the deck for the Myra run. The software uses the profile of the selected tube to determine the tip position for aspirating and dispensing, so any difference can cause incorrect pipetting behavior.

Incorrect plate height calibration.

It is essential that plate heights are correctly calibrated to avoid the tip colliding with the bottom of the well. For low-volume samples, this issue will impact aspiration, as the tip can become blocked at the bottom.

Excessive bubbles are present in the reagent tubes at the start of a run.
Excessive bubbles will affect level sensing. Such conditions can cause the Myra to aspirate from the bubble rather than the liquid, resulting in reduced volume during a transfer. Remove bubbles prior to the start of a run by spinning down the tubes.

What tubes can fit on the Myra?

Standard microtubes ranging from 0.1 to 5.0 mL, including screw-cap or flip-cap tubes, can be accommodated. Bottles up to 10 mL in size can also be used.

Longer tubes, such as blood collection vials, can be housed in the Myra9 and Myra Plus due to deep SBS sockets. However, the volume inside these tubes must be high enough for the pipette head and tip to reach it.

The clearance height for the standard Myra head with tip means that tubes above certain height limits cannot be used. For all Myra types, tubes above maximum height limits are incompatible.

Download the latest version of Workbench.

The latest version of Workbench software is 2.2.0

How BMS Support Can Help

Support
Packages

Submitting a Support Package ensures accurate and timely assistance. To generate this:

  • Open the Workbench software on the PC used that had the issue
  • Click the Help icon in the toolbar
  • Follow the prompts to generate a package

Upload the file. If the size exceeds the 256 MB upload limit, contact the support team, and an alternative method will be provided.

Assay Compatibility Issues

If you experience issues with assay compatibility, please provide:

  • A clear description of the issue
  • Any relevant Information for Use (IFU) documents associated with the assay or reagents

This allows the team to review assay conditions and provide targeted recommendations.

Warranty Claims and Repairs

If your instrument requires a warranty replacement or repair, contact BMS Support to discuss the issue and obtain a Return Merchandise Authorization (RMA) form.

Instruments must be returned clean and decontaminated. BMS reserves the right to refuse service on any equipment deemed unsafe due to contamination risk.

Mic Warranty Policy
Myra Warranty Policy

Support-Query

Support Query

Are you a current customer with a support query?
Please fill in the form below.

  • This field is for validation purposes and should be left unchanged.
  • This field is hidden when viewing the form
  • Drop files here or
    Max. file size: 256 MB, Max. files: 10.
      Upload instruction sheets or protocols for the application you wish to evaluate.