The bead-to-sample ratio is the variable that controls fragment size selection, and its precision is what separates a clean, tight Bioanalyzer trace from one that requires repeat preparation.
Automated SPRI bead cleanup and size selection
Solid Phase Reversible Immobilization bead cleanup is the step that sits between the enzymatic reaction and the next analytical stage in a molecular workflow.
Why bead ratio consistency is harder to achieve than it appears
SPRI bead cleanup follows a simple sequence. The bead ratio controls which fragments bind, then ethanol washes remove contaminants before elution releases the purified DNA. The bead-to-sample ratio is a volumetric relationship that is extremely sensitive to pipetting precision, particularly at the lower volumes used for single-sided cleanup and at the tighter ratios used for double-sided size selection.
A 5% error in a 50 µL bead transfer represents a deviation of just 2.5 µL, yet even small variations in bead-to-sample ratios can shift size-selection cut-offs and affect fragment size consistency between samples. When 24 samples are prepared by hand, uneven bead volumes across the plate are the primary source of sample-to-sample variability, leading to inconsistent fragment size profiles on the Bioanalyzer or Qsep.
Double-sided size selection compounds the error, because the left-side and right-side bead ratios must each be consistent. The effect on size cutoff accuracy is largest in NGS applications, where a tight insert size distribution is critical for sequencing efficiency.
Bead loss is the same risk here as it is in nucleic acid extraction. During supernatant removal after magnetic capture, a pipette tip that seals against the well base creates negative pressure. When the seal releases, the resulting suction can pull beads from the pellet.
Myra’s Dynamic Retract detects tip sealing and retracts incrementally before this pressure builds. This prevents bead loss at every wash and supernatant removal step, whether the cleanup runs on its own or as part of a library preparation workflow.
Where SPRI bead cleanup sits in molecular workflows
SPRI bead cleanup appears at several points across NGS, PCR, cloning, and diagnostic workflows.
NGS library preparation
Cleanup occurs after tagmentation, ligation, and PCR within library prep protocols. The uniformity of the fragment size distribution after bead cleanup directly affects cluster density and read quality, as well as the insert size profile in the final data. Myra runs all cleanup steps within a library prep run without a plate handoff.
PCR product purification
Removes excess primers, nucleotides, salts, and polymerase from PCR amplicons before sequencing, cloning, restriction digestion, or quantification. Single-sided cleanup at standard bead ratios suit amplicons above 100 bp, while size selection isolates specific fragment populations from complex PCR reactions.
Fragment size selection
Double-sided size selection narrows fragment size distributions before sequencing, particularly in PCR-free library preparation, where insert size depends entirely on bead cleanup rather than amplification. Left-side and right-side bead ratios independently define the lower and upper size cutoffs.
CRISPR-Cas9 workflows
Cleanup of donor DNA templates, purification of edited fragments before sequencing or analysis, and size selection of CRISPR components. Consistent cleanup at the editing verification step ensures that downstream sequencing data reliably distinguishes on-target and off-target events.
Fragment analysis
Preparation of size-selected fragment populations for capillary electrophoresis fragment analysis, microsatellite genotyping, and STR profiling. Consistent bead ratios determine peak resolution and prevent neighboring fragment classes from overlapping in the electropherogram.
Diagnostic sample preparation
Enrichment for target DNA or RNA fragments of specific sizes to improve the sensitivity and specificity of downstream diagnostic assays. Batch-to-batch variability in cleanup performance can affect assay performance thresholds, which makes consistent automated cleanup especially relevant in diagnostics.
Single-sided cleanup, double-sided size selection, and generic SPRI
The Myra liquid handling system supports all three principal SPRI workflow configurations. Bead volumes, ethanol wash volumes, air-dry timing, and elution volumes are all set through script parameters, so no modification is required for routine changes.
Left-side cleanup
Standard PCR product and library cleanup using a single bead ratio to remove fragments below the size cutoff. This cleanup step removes primers, primer dimers, adapter dimers, and small DNA fragments while retaining the target library or amplicon population. Bead-to-sample ratios of 0.6× to 1.8× are validated for input volumes up to 140 µL.
Size selection
Two sequential bead ratios define the lower and upper fragment size cutoffs. The right-side bead addition comes first to remove large fragments, followed by the left-side to capture the target population. Double-sided size selection is critical for PCR-free library preparation and any application requiring a narrow insert size distribution. Both bead volumes can be configured independently.
Custom SPRI protocols
Myra’s configurable SPRI script supports custom bead chemistries and non-standard protocols. Incubation times, wash steps, air-dry duration, and elution volumes are all configurable. Compatible with any paramagnetic bead-based cleanup chemistry that follows the standard bind-wash-elute format.
How Myra executes SPRI bead cleanup
The automated cleanup workflow follows the same structure for single-sided and double-sided configurations, with the double-sided protocol adding a right-side bead step before the left-side capture. All steps run on deck using the 96-well magnetic station.
Step 1. Bead addition and incubation
Step 2. Magnetic capture and supernatant removal
Step 3. Ethanol wash steps
Step 4. Air-dry
Step 5. Elution
What makes Myra suited to SPRI bead cleanup
Pressure-based liquid level sensing
MyraSense™ detects liquid below the dead volume of wells, which maximizes supernatant and wash removal. It does not rely on fixed-height aspiration, which may not reach actual liquid levels in partially filled wells.
96-well magnetic station
Integrated, automated magnet control keeps bead capture timing consistent across all samples in a run. Magnets are applied to groups of wells at defined intervals, which eliminates the variable capture timing that manual handling introduces across a plate.
Non-contact dispensing
Wash buffer is dispensed without tip contact with the destination surface. This protects pellet integrity during wash addition and reduces carryover risk between wells.
User-configurable parameters
Bead volume, ethanol volume, incubation times, air-dry duration, and elution volume are all set through parameters at the top of the script, so no editing is required to adapt the protocol to different bead ratios or sample volumes.
Intelligent mixing
Vortex, advanced, and simple mixing strategies are selected by volume and labware so the beads are fully resuspended before binding. Settled beads reduce the available surface area for DNA capture and shift the effective bead ratio, directly affecting size cutoff performance.
Dynamic Retract
Detects tip sealing against the well base during aspiration and retracts incrementally before negative pressure builds. This prevents bead loss from fluidic suction events during supernatant removal and elution transfer, protecting yield at every aspiration step over the pellet.
SPRI bead cleanup scripts for Myra
Scripts are validated by BMS, in partnership with kit manufacturers, or by the Myra user community. All kits follow the standard SPRI bind-wash-elute chemistry and are configurable within the same script framework.
| Kit | Verified By |
|---|---|
Zymo Select-a-Size DNA Clean & Concentrator MagBead Kit | BMS |
Wizard® MagneSil® Sequencing Reaction Clean Up System | Community |
MagSi-NGS PREP Plus Purification and Size Selection | BMSCommunity |
HigherPurity PCR Clean-Up | BMS |
SPRI Based Size Selection | BMS |
AmpliClean Cleanup Kit | BMSCommunity |
CleanNGS Double Size Selection | BMSCommunity |
CleanNGS Cleanup Kit | BMSCommunity |
AMPure XP DNA Cleanup | BMSCommunity |
Higher PCR product yield and tighter CV than manual preparation using CleanNGS beads
Labgene Scientific SA, Switzerland. Validated partner evaluation using CleanNGS beads on Myra.
Automated vs manual cleanup of PCR products across 16 samples
Labgene Scientific SA, a Swiss life science distributor and BMS partner, conducted a direct comparison between automated SPRI bead cleanup on Myra and manual preparation using CleanNGS beads (CleanNA). Sixteen PCR product samples were processed in parallel using both methods. Fragment size and size distribution were assessed by capillary electrophoresis using a BiOptic Qsep1-Plus fragment analyzer. The yield of purified PCR product was measured across all replicates to compare the mean yield and inter-sample consistency between the two preparation methods.
Significantly higher yield and improved consistency across all replicates
Automated SPRI bead cleanup on Myra produced significantly higher yields of purified PCR product than the manual method across all 16 samples. Replicate CV was also tighter for Myra-prepared samples than for the manual method, reflecting the consistency of automated bead ratio dispensing. Fragment size analysis by CE confirmed correct fragment sizes across all samples prepared by both methods, showing that automated cleanup on Myra preserved fragment integrity while improving yield and consistency.
Higher
PCR product yield vs manual method across all 16 samples
Tighter
CV across replicates compared to manual preparation
16/16
Correct fragment sizes by CE across all Myra-prepared samples
DNA yield using Myra vs. manual with CleanNGS beads.
The yield improvement is consistent with the bead ratio precision described above. Manual bead dispensing introduces well-to-well volume variability that shifts the effective bead-to-sample ratio, reducing the proportion of target fragments captured from each sample. Automated dispensing with consistent bead volumes across all wells maximizes capture efficiency while holding the size cutoff stable across the plate.
Bead cleanup in the context of the broader Myra workflow
SPRI bead cleanup on Myra connects directly to the upstream and downstream steps that surround it. It can run as a standalone cleanup or as an integrated step within a larger workflow on the same automated liquid handling platform.
qPCR setup
Eluates loaded directly as sample inputs. Reaction plates exported to Mic, QuantStudio, CFX, or Rotor-Gene Q.
NGS library preparation
Extracted gDNA or total RNA used as input for library prep across Illumina, ONT, and PacBio platforms.
Normalization and pooling
Post-quantification molar normalization, equimolar pooling, and Illumina loading preparation on the same platform.
Nucleic acid extraction
Post-extraction cleanup of eluates is performed before library prep or direct amplification.
Frequently Asked Questions
What bead-to-sample ratios can Myra run?
Can Myra do double-sided size selection, not just cleanup?
Do I have to edit the script to change bead volume or elution volume?
How does Myra prevent bead loss during supernatant removal?